Timer Trigger Powerdown Problems

The information provided is for help with Commtest Instruments products:



SYMPTOMS

When the instrument is set to:

We have found that some instruments will miss the scheduled wake up time; instead they wake up 24 hours later.

This only occurs:

CAUSE

The instrument uses a real-time clock to schedule the wake up time. Because of the 3 second warm up time (minimum) and a 2 second 'house keeping tasks' time, some instruments miss the scheduled wake up if the clock's second counter rolls over. This causes the next wake up time to occur in 24 hours. This only happens for PowerDown periods of fewer than 7 seconds.

SOLUTION

Do not enable PowerDown for Log Periods shorter than 10 seconds.

There are two methods that allow you to gather data every 5 seconds:

OR

No functionality is lost with these alternatives, which both avoid powering down for short times (i.e. offer no energy savings). In future releases the minimum PowerDown time will be set to 10 seconds.

FAQ ID: 11663 Last Reviewed: 27/09/2005