Incorrect Battery Charge Shown after Inserting a new Battery

The information provided is for help with Commtest Instruments products:



SYMPTOMS

When a new battery pack is inserted into the instrument the percentage charge and voltage values no longer match.

CAUSE

The vb instrument does not use intelligent batteries and so cannot identify the percentage of capacity of the new battery. However the actual voltage value displayed in the instrument will be accurate. The % value will be in error.

SOLUTION

To get the voltage value and the percentage value to realign you need to perform a battery recondition as outlined below.

Reconditioning a Battery

Inserting a Fully Charged Battery into the Instrument